Electric Techniques
Surface Potential Detection—Preparation
Rev. D Dimension 3100 Manual 299
Figure 16.5a Normal Jumper Configuration with Extender Electronics Module
Analog 2 Voltage Applied to the Tip (With Extender Electronics)
Notice that the jumper configuration in Figure 16.2k connects the Analog 2 signal from the
NanoScope III controller (± 12V range) to the tip, and is exactly the same as the jumper
configuration shown in Figure 16.2j, the standard configuration as shipped from the factory.
Remember to enable the Analog 2 voltage line as described in Section 16.2.2.
Figure 16.5b Jumper Configuration for Application of Voltage to Tip
ChuckGround/Bias
Analog 2 or
Unused
Auxiliary 2 Output
Ground
STM Signal
Indicates Jumpers
Gnd/OSC + DC signal
Unused
Ground
Ground
AFM Tip
Chuck
Sample
Tip
Chuck Ground/Bias
Unused
Auxiliary 2 Output
Ground
STM Signal
Indicates Jumpers
Unused
Ground
Ground
AFM Tip
Sample Chuck
Analog 2 or
Gnd/OSC + DC signal
Sample
Tip
Analog 2
Vista de página 318