Electric Techniques
Electric Force Microscopy
280 Dimension 3100 Manual Rev. D
Jumper Configurations Without Extender Electronics
As shipped from the factory, the jumper configuration on a Dimension Series system without the
Extender Electronics Module appears as shown in Figure 16.2e below. This configuration connects
both tip and sample to ground.
Figure 16.2e Normal Jumper Configuration
Sample Chuck Ground/Bias
Analog 2
Gain Select
Auxiliary D (to NanoScope III controller)
Ground
STM Signal (from Dimension head)
Indicates Jumpers
Unused
Ground
Ground
AFM Tip
Sample Chuck
Sample
Tip