Fluid Imaging
Sample Mounting
Rev. D Dimension 3100 Manual 153
10.2.3 Rubber Protective Skirt
A rubber protective skirt (see Figure 10.2b) is also provided with the fluid tip holder. This protective
skirt is used to protect the Dimension scanner from getting liquid on it. It is recommended that the
protective skirt be used with all wet samples. Skirts wear over time and may need to be replaced.
Figure 10.2b Rubber Protective Skirt
10.3 Sample Mounting
10.3.1 General Notes on Sample Binding
Immobilize samples for AFM imaging on a rigid support. Macroscopic samples (biomaterials,
crystals, polymer membranes, etc.) can be attached directly to a stainless steel sample disk with an
adhesive. Biological samples like cells, proteins, DNA, etc. are usually bound to a flat substrate
such as mica or glass. Many sample preparations have been developed and AFM applications
articles are an excellent source of information or sample binding.
Note: For a list of articles describing biological applications of AFM, including
sample preparation techniques, contact Veeco.
CAUTION: Do not attempt to operate the standard air tip holder in a fluid
environment. The standard tip holder has exposed electrical signal
lines that could short circuit if exposed to a conducting fluid.
ATTENTION: En milieu liquide, ne pas utiliser le support de pointe standard
prévu pour une utilisation à l’air. Le support de pointe standard
présente des contacts électriques qui peuvent être endommagés
(court-circuit) suite à une exposition à un liquide conducteur
d’électricité.
VORSICHT: Versuchen Sie nicht, den Standard-Luft-Cantileverhalter in
Flüssigkeiten zu betreiben. Am Standard-Cantileverhalter liegen
elektrische Leitungen frei, die in leitfähigen Flüssigkeiten
kurzgeschlossen werden könnten.