Fluid Imaging
Operating Principles
168 Dimension 3100 Manual Rev. D
10. Adjust the Drive amplitude until obtaining an RMS Amplitude of 0.3 - 0.6V. This value
has produced good results for protein samples like RNA polymerase and lysozyme.
Note: In general, larger RMS amplitudes (approximately 2V) work better for taller
samples such as cells.
Figure 10.5e Cantilever Tune Frequency Sweep
11. Click OK. The parameters set in the Cantilever Tune control panel appear in the Realtime
control panel.
12. Click on CANCEL to exit the Cantilever Tune command and leave the parameters
unchanged.
Cantilever
Response
0.25 V/div
Center Frequency — 18.44 KHz
0.50 KHz/div
Frequency Sweep
Setpoint