Dell 3100 Especificações Página 162

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 161
TappingMode AFM
Withdraw the Tip
142 Dimension 3100 Manual Rev. D
9.2.11 Optimize Scan Parameters
The procedure to optimize the scan parameters is similar to that of Contact Mode.
1. In the Scan Controls panel, set Slow scan axis to Disabled.
2. Select View > Scope Mode.
3. Observe the agreement between the trace and retrace lines. Adjust the Scan rate, Gains, and
Amplitude setpoint to bring these two lines in to coincidence. A more detailed discussion of
these parameters can be found in Section 9.4.
9.3 Withdraw the Tip
1. Select Withdraw from the Motor menu. The SPM stops scanning, and ascends to the
sample clearance height defined in the SPM Parameters menu.
2. Select the Stage > Load New Sample option to replace or move the sample.
3. Use the Focus Surface command to move the SPM up if you desire more clearance between
the tip and sample.
CAUTION: Never withdraw samples without verifying that the tip has
adequate clearance during the entire sample removal sequence.
Vista de página 161
1 2 ... 157 158 159 160 161 162 163 164 165 166 167 ... 375 376

Comentários a estes Manuais

Sem comentários