
Scanning Tunneling Microscopy (STM)
Troubleshooting Operation of STM
190 Dimension 3100 Manual Rev. D
resistance measurements. However, oxide and contamination layers on the sample surface can
make imaging very difficult. Higher bias voltages are required for these types of samples.
Problems During Realtime Operation
This section deals with problems that are associated with the Realtime aspects of the NanoScope
software although they often point to the scanning head or microscope.
Image goes out of Range
If the Z center position tends to go out of range at the edges, top and bottom or right and left of the
scan, then try the following:
•Level Sample to Head—The Z center position is affected by sample tilt and the
Dimension head may be less tolerant of sample tilt due to its small vertical range
compared to the large, lateral scan range. Try leveling the head relative to the sample by
noting the trend in Z center position and adjusting the head tilt accordingly. The View >
Scope Mode with the Show slope option can also be very helpful in eliminating tilt
along the scan direction.
•Measure Coupling— A broken wire to one of the electrodes can cause this problem and
will usually result in a large, monotonic increase (or decrease) in Z center position from
the bottom of the scan to the top.
Z drift
If the Z center voltage tends to drift out of range rather quickly after the head is engaged, it may be
a result of the following:
1. Thermal Drift— Allow some time for the temperature to stabilize if the microscope and/or
scan head has been stored in a cold place overnight. Drift can be minimized by keeping the
STM in a thermally stable environment.
2. Tip not Tight—Make sure the tip is held tightly in the tip holder. Push and pull on the tip
with the tweezers to see that it is not loose. Increase the bend in the tip if the tip seems loose.
3. Sample Hold-Down—Verify that the sample is flat against the chuck and cannot move
vertically. This may be become problematic if the sample has been affixed using uncured,
conductive epoxy, and/or if foam adhesives are used.
Realtime Image Hides Features
If the Realtime image appears to be flat, but captured images reveal detail, then try rotating the
sample 90° in the microscope. The Realtime image leveling software tends to hide features that are
parallel to the X-scan direction.
Comentários a estes Manuais