Scanning Tunneling Microscopy (STM)
STM-Specific Information and Operations
186 Dimension 3100 Manual Rev. D
Vibration Isolation
The microscope should be isolated from sources of vibration in the acoustic and subacoustic
frequencies. This requirement can be relaxed somewhat for large-scale images, but atomic-scale
work is extraordinarily sensitive to ordinary room vibrations.
As a final note, the best way to reduce coupling from vibrations is to eliminate as many sources of
vibration as possible. Remember that vibrations can be transmitted to the microscope over the
cable. To reduce this phenomenon, prevent tension in the cable and keep it away from fans and
other noise sources. Also, keep the microscope away from sources of acoustic noise. Loud
conversation can disrupt atomic scale images. Air currents can also disturb atomic images, so it is
best to run the microscope with the cover on.
Piezoelectric Sensitivity and Derating
The sensitivity of the piezoelectric tube in nanometers per volt and the derating of the piezo
sensitivity as the scan voltage is decreased are calibrated using the Autocalibration command in
the Offline > Utility menu. Prior to using the Autocalibration command, the calibration images
must be captured with the Capture Calibration command in the Realtime > Capture menu.
Generally, a silicon calibration standard is used for the “G” scanner-based, Dimension head.
The sensitivity of the piezoelectric tube decreases as the applied range of the applied scan voltage is
reduced. Over a small scan range, the sensitivity of the piezo is constant and a plot of scan size
versus scan voltage is linear. However, for large scans, sensitivity variations are an important factor.
For example, the G head shows sensitivity variation. A plot of the sensitivity of a long-range head
as a function of scan voltage looks similar to the graph in Figure 11.3b. As shown, the Derating
(the slope of the head's sensitivity plot) indicates how much the piezo sensitivity decreases as the
scan voltage is reduced.
Figure 11.3b Scan Derating Graph
The NanoScope software accounts for the reduction in sensitivity as the scan voltage is decreased
through the Derating parameters in the Microscope > Calibrate > Scanner dialog box. As
described in the following paragraphs, the Head Autocalibrate command in the Offline > Utility
menu can be used to calculate the sensitivity and derating parameters of STM heads.
Scan Size (V)
110 220 440
Sensitivity (Å/V)
slope = Derating
Maximum Sensitivity
300
250